Tài liệu Follow-Up of the patient who has received pacemaker pptx

27 400 0
Tài liệu Follow-Up of the patient who has received pacemaker pptx

Đang tải... (xem toàn văn)

Tài liệu hạn chế xem trước, để xem đầy đủ mời bạn chọn Tải xuống

Thông tin tài liệu

Authors: Moses, H Weston; Mullin, James C Title: A Practical Guide to Cardiac Pacing, 6th Edition Copyright ©2007 Lippincott Williams & Wilkins > Table of Contents > 11 - Follow-Up of the Patient who has Received a Pacemaker 11 Follow-Up of the Patient who has Received a Pacemaker Initial Pacemaker Clinic Follow-Up Visit An organized system of follow-up is important for all device patients and is best facilitated through the patient's membership in a formal pacemaker/ICD follow-up clinic The first formal clinic follow-up visit is usually scheduled at approximately months after implantation to allow for completion of the lead-tissue maturation process, which may affect sensing and capture thresholds Reprogramming, to account for any threshold change since implant, and still provide a safety margin along with improved battery life, may be done at that time Optimization of device function can be performed noninvasively The patient generally will have additional questions at this time, and the pocket can be evaluated for appropriate healing Focused History and Physical The history should focus on symptoms related to impaired cardiac output that may reflect potential pacemaker malfunction These include palpitations, dizziness, presyncope/syncope, or any symptom that may resemble those experienced preimplantation The possibility of pacemaker syndrome always should be considered as a potential cause for recurrent or new symptoms Complaints related to potential pacemaker infection also should be explored, including fever, chills, recurrent respiratory illness, or reports of swelling, drainage, or tenderness in the region of the pocket Table 11-1 lists some symptoms that should lead to the suspicion of potential problems Fortunately, most of these problems are quite rare The examination consists of measuring vital signs and inspecting the pocket area Attention should be paid to the presence of localized tenderness, swelling, or redness The patient may complain of symptoms related to progression of his or her underlying cardiac disease so that the follow-up visit presents an opportunity for reevaluation of the patient's general cardiac status P.158 TABLE 11-1 Symptoms that may alert the clinician to pacemaker problems Patient symptoms Consider I have a constant hiccup Diaphragmatic pacing My shoulder jumps Pectoral muscle pacing My pacemaker is sore Pacemaker pocket infection I am sick Pacemaker lead infection My arm hurts Brachial plexus irritation My heart is racing Pacemaker-mediated tachycardia or other tachycardias unrelated to the pacemaker I am dizzy/I passed out Loss of capture or inappropriate bradycardia My pacemaker has moved and is in my Pacemaker migration armpit, lower chest, etc My hand (or arm) is swollen Clot in the vein I have prominent veins on the shoulder Clot in the vein nearest the device Electrocardiogram Twelve-lead EKG's are no longer really required on a first visit since excellent information is obtained from interrogating the device itself If an EKG or rhythm strip is performed, it is important to note that a unipolar spike can distort the baseline to a degree that can be mistaken for true capture This is illustrated in Figure 11-1 Bipolar spikes tend to be smaller than unipolar spikes (because the electricity is traveling the shorter distance through the body) and may be harder to identify If the patient is in a clinic, a look at various leads of the electrocardiogram (ECG) may clarify the presence or absence of capture, in the case of confusing rhythm strips received over the telephone Fusion and pseudofusion beats are shown in Figure 11-2 A fusion beat is a QRS complex that has been formed by depolarization of the myocardium, initiated by both the pacemaker spike and the patient's intrinsic beat A pseudofusion beat is a QRS complex that is formed by a depolarization of the myocardium initiated by a patient's intrinsic heartbeat completely; however, a pacemaker spike is present, distorting the QRS complex These can be normal occurrences in pacemaker patients due to timing of the depolarization wave in the ventricle A rarely used term is pseudopseudofusion beat This would be an atrial spike that, by chance, lands at the beginning of a QRS complex thus mimicking a pseudo-effusion beat P.159 Figure 11-1 Unipolar Pacemaker Rhythm Strip Top: A unipolar pacemaker with a large pacemaker spike and distorted baseline due to afterpotential There is no capture of the heart during this time The distorted baseline can be mistaken for a QRS complex Bottom: A rhythm strip in the same patient with repositioning of the lead and capture of the patient's heart Note the appearance of a T wave, which is often helpful in identifying true capture in this situation Magnet Rate To obtain the magnet rate, a standard magnet is applied over the pocket area for a brief period The presence of the magnetic field causes a switch to temporarily convert the pacemaker into the VOO (asynchronous pacing) or P.160 DOO mode in the case of a DDD device The rate at which the pacemaker fires during the magnet application depends on the model of the pacemaker Some models fire at a slower rate than the programmed rate, whereas others fire at the same or at a faster rate A common question raised by clinicians concerns the safety of using the magnet mode because, theoretically, a pacing spike occurring on the T wave could induce ventricular arrhythmias This is rarely a practical problem, however, unless significant cardiac ischemia is present The routine use of magnet mode during transtelephonic evaluations has been shown to be generally quite safe Figure 11-2 Fusion and Pseudofusion Beats Intrinsic beats and paced beats are shown A fusion beat is a combination of the patient's intrinsic beat and a paced beat A pseudofusion beat is a QRS complex caused by an intrinsic beat that is distorted by the patient's spike, which does not obviously depolarize the myocardium These findings are normal and not represent pacemaker malfunction Pacemaker Interrogation The device then should be interrogated to establish the programmed parameters The interrogation also should include evaluation of variables referred to as measured data, which include lead information (atrial and/or ventricular), such as pulse amplitude (volts), pulse width (milliseconds), current, impedance, and thresholds Data can be printed out on paper, but a more common trend is to interact with electronic medical records and place data directly into the electronic record This is a complex area, but a commonly used term is HL7 compatibility This stands for Health Level compatibility and involves a complex set of standards (the vast majority of which are totally unrelated to the pacing industry) for integrating hospital data into physicians' electronic medical records Establishing the Underlying Rhythm An important step is to establish the patient's underlying rhythm and to determine whether the individual is pacemaker dependent, a condition in which hemodynamic instability would result on abrupt cessation of pacing Because of the increased risk of serious consequences that patients may experience in the event of pacemaker malfunction, the chart should be labeled clearly in some manner (a brightly colored sticker) that the patient is pacemaker dependent The underlying rhythm can be uncovered by reprogramming to VVI at a low rate such as 30 to 40 bpm and recording the underlying rhythm It is often helpful to decrease the pacemaker rate gradually to allow the native automatic focus to “warm-up.”However, an abrupt reduction in the paced rate will more accurately simulate the response to transient pacemaker malfunction Evaluation of Sensing Threshold The determination of sensing threshold establishes the adequacy of the sensed amplitude of the R wave (and/or P wave) Sensitivity is a measure P.161 of how easily the pacemaker can “see”the R wave or P wave Sensitivity values, expressed in millivolts (mV) may be selected with the programmer and range from less than 0.5 mV to greater than mV The greatest sensitivity is associated with the smallest numeric value in mV, whereas the least sensitivity is associated with the largest mV value Each value represents a cutoff for the ability of the pacemaker to detect the height of a given intracardiac electrogram For example, an R wave with an amplitude of 2.4 mV will be just missed with a sensitivity setting of 2.5 mV/cm By changing the sensitivity to 1.0 mV/cm, the generator will sense any signal with an amplitude greater than 1.0 mV and be able to track the R wave of 2.4 mV To establish the sensing threshold, sensitivity is decreased progressively until the sensed signal is lost The point at which that occurs is referred to as the sensing threshold Because the concept of sensitivity is difficult to grasp initially, an analogy is often helpful Consider sensitivity value to refer to the height of a fence that blocks the view of objects (an R wave or P wave) on the other side Although the heights of the objects are fixed, the height of the fence can be varied By lowering the fence (from 4.0 to 0.5 mV/cm), the sensitivity is increased and the object can be brought into view (sensed); by raising the fence, the sensitivity is lowered and the signal can be shielded from view (not sensed) In VVI (or AAI) situations, the lower rate limit of the pacemaker is programmed to a value lower than the intrinsic rate before testing is begun because you want to be able to see the native activity If the patient's native ventricular rate is too low, the sensing threshold test cannot be carried out If you begin with the most sensitive level (lowest number) and progressively decrease sensitivity, threshold is reached when pacing is first observed to occur (i.e., the native rhythm is no longer “seen”) In DDD applications, the atrial sensing threshold is tested by performing the same process: Threshold is reached when atrial pacing is first noted To test the ventricular sensing threshold, the atrioventricular (AV) interval is lengthened to exceed the PR interval in order to allow the native QRS to be seen The ventricular sensitivity then is progressively decreased from its most sensitive value until a ventricular pacing artifact begins to appear at the end of the expected AV interval Because of the large R wave amplitude in most patients, R wave sensing will still occur even at the least sensitive setting, in which case the sensing threshold is recorded as being greater than this (e.g., >4.0 mV/cm) Typically, you will program in the sensitivity value that is closest to one-half the sensing threshold value Evaluation of Capture Threshold Capture threshold refers to the lowest pulse-output value associated with stable capture of the myocardium and thus allows determination of the lowest safe pulse output associated with greatest battery longevity In contrast to P.162 sensing threshold testing, the lower rate limit of the pacemaker is reprogrammed to a value greater than the intrinsic rate before testing is begun As in sensing threshold determination, most devices feature an automatic capture threshold test Frequently, the pulse width is the variable that is decreased progressively until the loss of capture, although some manufacturers have automatic threshold determinations in which the pulse output is progressively decreased at a fixed pulse width Pulse amplitude and pulse duration can be compared by noting that a doubling of pulse output will quadruple the delivered energy, in contrast to only doubling delivered energy when pulse duration is doubled In VVI applications, the ventricular capture threshold can be determined by progressively decreasing the pulse output in the graded settings or by progressively decreasing the pulse width until loss of capture In DDD applications, the method of capture threshold testing depends on the state of underlying AV conduction With intact AV nodal conduction, the atrial capture threshold test is carried out as follows: The device is programmed to the DDD mode with a lower rate limit set to about 20 bpm greater than the native sinus rate The AV interval is prolonged if “native”conducted QRS complexes are not noted The atrial pulse output is decreased in a stepwise manner until the loss of atrial capture occurs, which is signaled by the appearance of ventricular pacing at the programmed AV interval The ventricular capture threshold is determined by programming the device in the same manner as for atrial capture testing but with the AV interval shortened to less than the PR interval The ventricular pulse output then is decreased in a stepwise fashion until the loss of ventricular capture occurs This will be manifest by the appearance of native QRS complexes In situations where AV conduction is impaired, such as second- or third-degree AV block, atrial and ventricular capture threshold testing is carried out in a slightly different manner For atrial capture testing, the device is programmed to the DDD mode with the lower rate limit to 20 bpm greater than the sinus rate and with the shortest possible postventricular atrial refractory period (PVARP) The atrial output then is decreased in a stepwise manner until the observed occurrence of an irregular rhythm as a result of the occurrence of mixed AV synchronous and AV sequential pacing Ventricular capture threshold testing is carried out by programming the device to DDD with the lower rate limit to 20 bpm greater than the native ventricular rate The ventricular pulse output is decreased in a stepwise fashion until a loss of ventricular capture and heart rate slowing are noted For situations in which the native ventricular rate is too slow or an underlying ventricular rhythm is not evident, the emergency VVI feature that most programmers have allows for rescue pacing and allows time to reprogram the ventricular pulse output to an appropriate value Because of the time required to attain a stable capture threshold following implant, you should probably not attempt to decrease either the atrial or ventricular voltage output until the initial follow-up visit at months At that P.163 time, usually the pulse output value is reprogrammed to at least twice the capture threshold value Tables 11-2 and 11-3 summarize the steps necessary to perform both sensing and capture threshold testing TABLE 11-2 Sensing threshold testing VVI Program LRL to 20 bpm less than native ventricular rate and with highest ventricular output Stepwise decrease ventricular sensitivity (low to high mV/cm) Threshold moment when ventricular pacing begins DDD Atrial Program DDD with LRL 20 bpm less than sinus rate and with highest atrial output Stepwise decrease atrial sensitivity (low to high mV/cm) Threshold moment when atrial pacing begins Ventricular Program DDD with LRL 20 bpm less than active ventricular rate, longest available AV, and highest ventricular output Stepwise decrease ventricular sensitivity Threshold = moment when ventricular pacing artifact can be seen at end of QRS or in ST segment AV, atrioventricular; LRL, lower rate limit Sensor Evaluation In patients with pacemakers that have programmable rate modulation, you may need to reprogram the variable pertaining to this feature Presently, the available sensors include activity and minute ventilation or combinations of these The manner in which the sensor is initially set up depends on the manufacturer, so you should refer to the manual to obtain guidance Generally, ascertain whether sinus node dysfunction is present at the time of implant, and this becomes the primary criterion for selecting this type of device Always remember that patients who are taking medications that have negative chronotropic effects often exhibit sinus node dysfunction that may benefit from the DDDR device If chronotropic incompetence is suspected but is minimally present, the device can be programmed to the DDD mode at the time of implant When the patient returns for the 3month follow-up visit, an inquiry into tolerance of mild to modest physical activity may reveal symptoms suggesting inadequate pacemaker rate response and prompt reprogramming of the DDDR mode Modern pacemakers generally allow a histogram demonstrating the distribution of rates over a long period of time Obviously, if the patient has a heart P.164 rate at the lower programmed rate, then there is strong evidence of inadequate pacemaker rate responsiveness To verify this, it is useful to have the patient walk two flights down a stairwell and back, and then reassess the heart rate response on return to the examining room This can be done with the rhythm strip or through telemetry Changes in the sensor-driven upper rate limit can then be made based on clinical judgment TABLE 11-3 Capture threshold testing VVI Program LRL to 20 bpm greater than native ventricular rate Stepwise decrease in pulse output Threshold = loss of ventricular capture and sudden rate slowing DDD (with intact AV conduction) Atrial Program DDD with LRL to 20 bpm greater than sinus rate Stepwise decrease in atrial pulse output Threshold = appearance of ventricular pacing Ventricular Program DDD with LRL to 20 bpm greater than sinus rate and short AVP (AVI

Ngày đăng: 24/01/2014, 01:20

Từ khóa liên quan

Tài liệu cùng người dùng

  • Đang cập nhật ...

Tài liệu liên quan